BCS-975R Premium Battery Cyclers - BioLogic
Premium Battery Cyclers

BCS-975R Premium Battery Cyclers.

High power, ultra-precise battery cyclers with fully native EIS capability

The BCS-975R is a high-power, regenerative battery cycler. With 8 channels, each capable of 75 A and up to 300 A in parallel, it is purpose built for pouch and prismatic cells.

Request a quote Add to selection Support page

BioLogic’s Premium Battery Cyclers: High-end cyclers for demanding needs.

Our Premium Battery Cyclers are composed of the BCS-900 Series. This system merges unmatched performance with a user-centric design, owing to the groundbreaking BT-Lab® Suite, your trusted partner in battery testing and data analysis software.  This series of ultra-precise cyclers features five current ranges, fast bandwidths (rise and fall times), smart sampling, interfacing with external equipment and temperature probes, amongst many other high-end functions.
 
The result is a powerful, modular system suitable for any organization specializing in battery testing and research, from industry to academia, quality control to certification.

The BCS-975R Battery Cycler: High-Power Testing for Advanced Research Applications.

The BCS-975R is a regenerative battery cycler designed for pouch and prismatic cell testing and cycling. With native EIS and extendable current up to 300 A, this instrument is designed for high-power performance battery testing. Built with uncompromising performance in mind, the BCS-975R delivers exceptional precision and versatility through its modular design, including:

  • 8 channels maximizing your testing capacity
  • Up to 75 A per channel
  • 3 current ranges for precise control
  • Optional Auxiliary temperature monitoring
  • BT-Lab® Suite Software (including BT-Test™ & BT-Analysis™)

Video

Precision, accuracy & modularity: three reasons to choose the BCS-900 battery cycler series.

Three current ranges yield excellent accuracy allowing to adapt to the various current level situations to which the cell tester is confronted to with no compromise on accuracy.
 
With a 24-bit current and voltage converter, the BCS-975R battery cycler series has been developed to measure your high-power test results with high resolution.
 
An optional built-in FRA (Frequency Response Analyzer) is available for every BCS-900 module enabling native and fully integrated Electrochemical Impedance Spectroscopy (EIS) experiments over the frequency range of 2 kHz to 10 mHz on each channel in either galvano or potentio mode. EIS measurements are as fast and precise as those obtained with our research-grade equipment – no compromise on quality has been made. With EIS, the evolution of your batteries’ parameters can be precisely determined (SOC, SOH, internal resistance Ri). Want to know more about Electrochemical Impedance Spectroscopy and why it is use in for battery testing? Click here.
 
The instrument offers a fast data acquisition with a 1 ms sampling rate. Learn more about data of interest to work more efficiently.

BT-Lab® Suite: Comprehensive from reliable tests to efficient analysis

The BCS-900 series Battery Cyclers feature BT-Lab® software Suite, the latest advancement in battery testing technology. This groundbreaking software integrates state-of-the-art features to enhance control, monitoring, and analysis capabilities for rigorous battery testing applications.
Ensuring the utmost reliability in cycling procedures, continuous monitoring throughout testing is provided by our embedded BT-Test™, guaranteeing consistent and accurate results. Efficiency in data analysis is further optimized through the advanced BT-Analysis™ software, enabling rapid and insightful interpretation of complex datasets. Specifically engineered for battery testing, essential features such as native EIS provide researchers and engineers comprehensive tools to delve deeper into battery performance and behavior.

BT-Test™: Superior performance in battery testing software.


BT-Test™ integrates cutting-edge features that empower researchers with unparalleled control and flexibility in designing, running and monitoring cycling experiments. Researchers can effortlessly control variables to tailor experiments to their exact specifications, ensuring the acquisition of highly accurate and insightful data.

BT-Analysis™: Maximum efficiency with automated batch data processing.


For demanding battery data processing, in-depth and automated analysis is completed with BT-Analysis™. Analyses adapt to specific needs with unmatched flexibility. Access to critical data anytime, anywhere – online or offline – ensures key data is never missed. BT-Analysis™ empowers detailed analysis by displaying data in various formats, including vs. cycles, loops, and steps. It even displays application-oriented graphs, tables, and statistical tools in order to gain a deeper understanding of battery behavior.

Work anywhere

The BT-Lab® Software Suite offers an unprecedented amount of flexibility, giving researchers more freedom to access their data. Using BT-Test™, Test Run monitoring can be as simple as opening a web browser. With BT-Analysis™, live data management and processing can be accessed remotely.

 

 

Specifications

 

BCS-975R battery cycler
Channels per module8
Voltage
Range0 V to 5 V
Control resolution 100 µV
Measurement resolution0.360 µV (24 bit)
Accuracy15 mV + 0.05% of setting
Slew rate2.5 kV/s
Current
Max (continuous) per channel±75 A;
up to ±300 A, with 4 channels in parallel
Ranges3: 75 A down to 100 mA
Control resolutionDown to 360 nA
Measurement resolutionDown to 8.9 nA (24 bit)
Accuracy 0.03% of FSR* + 0.03% of setting
EIS
Built-inOptional on each module (multiplexed across 8 channels)
Range2 kHz – 10 mHz
Measurement
Acquisition time1 ms
Time base5 ms
Additional measurement
ThermocoupleOptional Auxiliary
Cell connection
4 terminal leads
General
Height16U
Weight120 kg
Power consumption5000 W

*FSR: Full Scale Range

What should I do if I have an issue?

If you have any questions or if any problem occurs that is not mentioned in this website, please contact your local retailer.  Your provider’s highly-qualified staff will be glad to help you.

Should you need to contact us to communicate an error, we will need to have specific information concerning your hardware and software configuration.

Please keep information on the following at hand:

  • Description of the error (the error message, mpr file, picture of setting or any other useful information) and of the context in which the error occurred. Try to remember all the steps you had performed immediately before the error occurred. The more information on the actual situation you can provide, the easier it is to track the problem.
  • The serial number of the device located on the rear panel device.
  • The software and hardware version you are currently using. On the Help menu, click About. The displayed dialog box shows the version numbers.
  • The operating system on the connected computer.

Where can I download the software to control my instruments?

To download BT-Lab® Software Suite, please consult the software support page:

BT-Lab® Software Suite

Accessories

You may want to complete your set-up with.

The following equipment may help you with your research. A full range of battery sample holders are available from BioLogic. They are compatible with all our battery cyclers. See a selection below and go and visit the BioLogic accessory pages to find out more.

Greater Scope. Increased flexibility.

Related topics